Operation 3-9
January 2006
Factory Default Settings As shipped from the factory, the Optica is normally programmed with
the configuration shown in Table 3-2 below.
Table 3-2: Factory Default Values
Function Setting
Analog Output A
Analog Output B
Humidity, Units: Tdew °C, Range: –40°C to +60°C
Temperature, Units: Tmp °C, Range: 0 to 100°C
Pressure Input 1
Units: psia, Range: 0–30, Default pressure: 14.70,
Process pressure: disabled, 14.70
Alarm 1
Alarm 2
Disabled, Humidity, Units: Tdew °C, Set Point: 100.00
Disabled, Temperature, Units: Tmp °C, Set Point: 100.00
Auto Balance Interval: 720 minutes, Automatic, Disabled, Hold
Data Field 1
Data Field 2
Data Field 3
Data Field 4
Data Field 5
Data Field 6
Humidity, Units: Tdew °C, Range: –40.000 to +60.000°C, Color: green, 1 decimal
Humidity, Units: %RH, Range: 0 to 100, Color: blue, 1 decimal
Temperature, Units: Tmp °C, Range: 0 to 100, Color: red, 1 decimal
Pressure, Units: psia, Range: 0 to 100, Color: yellow, 1 decimal
Humidity, Units: ppmv, Range: 0 to 1,000,000, Color: brown, 1 decimal
Humidity, Units: ppmw, Range: 0 to 100, Color: violet, 1 decimal
Buzzer 15 msec
General Data Fields: 3, Lockout: disabled, Offsets and filters: 0
Special Molecular Weight of Gas: 28.9645
User Equation None entered
Communication
Parameters
Mode: All, Baud rate: 9600, Parity: none, Data bits: 8, Stop bits: 1,
Flow control: none
Serial Outputs
Humidity: Units: Tdew °C, Format: DP=##.#, Separator: CR-LF, Terminator: CR-LF,
Interval: 1 second, Time Stamp: Enabled, Show Status: Enabled
Temperature: Units: Tmp °C, Format: DP=##.#, Separator: CR-LF, Terminator: CR-LF,
Interval: 1 second, Time Stamp: Enabled, Show Status: Enabled
Pressure: Units: psia, Format: DP=##.#, Separator: CR-LF, Terminator: CR-LF,
Interval: 1 second, Time Stamp: Enabled, Show Status: Enabled
Selected
Outputs
Humidity, Units: Tdew °C
Humidity, Units: % RH
Temperature, Units: Tmp °C
Data Log
Status: Disabled, Interval: 1 second, Separator: comma, Terminator: CR-LF,
Parameters: humidity, Units: Tdew °C