Programming the VGA Optica 4-13
January 2006
Datalog Enter parameters for automatically logging data within the Optica.
This menu is accessed by selecting the Data Log control button on
Menu 1 and pressing ENTER.
Figure 4-5: Datalog Menu
The large window below the center of the screen shows a list of
parameters chosen for logging.
Table 4-10: Datalog Options
Item Function Available Options
Status* Enable or disable datalog. Enabled/Disabled
Interval* Enter the logging interval in seconds. Enter a number.
Decimals Enter the number of decimal places for logged data. Enter a number (6 maximum).
Separator* Choose separator to be used between parameters. Space, Comma, Tab
Terminator* Choose the terminator(s) for each group of data. CR, CR-LF, LF
Parameters Set the parameter to be programmed (below). Humidity, Temperature,
Pressure, or User Equation
Units Set the units for the selected parameter. See Table 4-1 on page 4-4.
REMOVE Control button to remove selected item from the list. Select button and press ENTER.
ADD Control button to add the selected unit to the list. Select button and press ENTER.
RESET Delete datalog files. Select button and press ENTER.
DOWNLOAD Control button to display the logged data on the
Download screen (see below).
Select button and press ENTER.
*Indicated items apply to the entire datalog file.