January 2006
5-8 Programming the 4x40 Optica
Automatic Cleaning and
Balance Function
Note: For Optica Analyzers with earlier versions of software, see
Appendix F.
Optica Analyzers with version XXX software have been upgraded to
include the ability to program the PACER self-cleaning and
rebalancing cycle to run once per day at a preset time. This is referred
to as a Clock Time Interval. When the unit is programmed in this
manner, the front panel PACER softkey is disabled. The Elapsed Time
Interval initiates the PACER at a preset time after the last PACER was
run. The automatic balance cycle will always run upon power up of
the analyzers.
To program the Automatic Cleaning & Balance Function on the
Optica 4X40 monitor:
1. Make sure the internal clock of the Optica has the correct time.
From the main screen:
a. Press the ENTER key.
b. Press the DOWN soft key until Set Time and Date is
displayed.
c. Press the >>>> soft key.
d. Press the >>>> soft key.
e. Highlight the Date and Time fields using the tab key. Use the
keypad and softkeys to set the correct time.