Programming the 4x40 Optica 5-9
January 2006
Automatic Cleaning and
Balance Function (cont.)
2. Set the Time Programmed Balance Cycle:
a. Press the ENTER key to access the Main menu.
b. Press the DOWN soft key until Auto Balance is shown.
c. Press the >>>> soft key.
d. Press the >>>> to set the interval. Use the soft keys and the key
pad.
e. Press the ENTER key when done.
f. Press the DOWN soft key to select the pacer type.
g. Press the >>>> to enter selection mode.
h. Press the DOWN soft key until the desired pacer type is
displayed.
i. Press the ENTER key to select.
j. Press the DOWN soft key to enable the PACER status.
k. Press the >>>> soft key to enter selection mode.
l. Press the UP or DOWN soft key until the enabled status is
displayed.
m. Press the ENTER key to save the selection.
n. Press the <<<< key several times to exit the menu.
o. Under the Type pull down menu there will be four selections
(see Table 5-9 below).
Note: The PACER function will provide more thorough cleaning than
the AUTO function because it first develops a thick dew/frost
layer into which soluble contaminants dissolve. When heated,
some of the contaminants are flash evaporated and the
remaining residue accumulates in clusters, resulting in the
cleaning of much of the mirrors surface. The PACER cycle
generally takes longer to complete.
Table 5-9: Automatic Cleaning and Balance Functions
Item Function Type
AUTO Heats the mirror and balances the optics. Elapsed Time Interval
PACER
First cools the mirror to develop a thick dew/frost
layer, then heats and balances the optics.
Elapsed Time Interval
AUTO-D Heats the mirror and balances the optics. Clock Time Interval
PACER-D
First cools the mirror to develop a thick dew/frost
layer, then heats and balances the optics.
Clock Time Interval