EasyManua.ls Logo

GE Optica - Page 51

GE Optica
150 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation 3-7
January 2006
Measuring Vapor Pressure
With a Pressure Sensor
To measure vapor pressure with a 4-20 mA, 0-30 psia pressure sensor,
connect the equipment as shown above and make the following
entries in the Pressure Input menu section.
Input: Iin 4-20
Units: psia
Upper: 30.00
Lower: 0.00
In addition, to accommodate the pressure expansion, the following
entries are required in the Process section:
Status: enabled
Pressure: 500 (specify the process pressure with the same units
as the default pressure specified above)
Measuring Dew Point With
or Without a Pressure
Sensor
In Scenario 2, the dew point measurement requires both the sensor
pressure and the process pressure to be known. Sensor pressure can
be entered as a default value, or measured, as above, and the process
pressure must be entered in the Process Pressure field.
Example (using standard atmospheric conditions at 25°C)
P1 = 500 psia
P2 = 14.7 psia
As measured by the GEI chilled mirror dew point sensor:
Tdew2 = –40°C @ P2
e2 = 0.1283 mbar
Per Dalton’s Law of Partial Pressure:
e1 = (P2/P1) × e2 = (500/14.7) × 0.1283 = 4.36 mbar
Using standard vapor pressure equations, the actual pressurized dew
point is calculated by the Optica to be:
Tdew1 = –4.022°C

Table of Contents

Related product manuals