Page #22 background image
Description
10
Service instructions
5015108-00 Edition 0403
V1513
1 Pre-rinse 1
2 Pre-rinse 2
3 Wash
4 Neutralize
5 Post-rinse 1
6 Post-rinse 2
7 Final rinse
8 Disinfection
9 Drying
Program phases
A Alkaline detergent
B Neutralisation
C Instrument milk (extra
equipment)
If instrument milk is
does, neutralisation is
not dosed.
Abort at start of process
A started program can be aborted within ten seconds of the door locking. To abort a started
process, press V. During the period when the program can be aborted, a yellow lamp flashes at
M. The door is unlocked and opened automatically and the machine can be restarted in the
usual way.
Aborting an ongoing process
During a process, the machine can be stopped with the main switch. If the power is switched on
within 60 seconds, the program continues until the end. If the power is switched on after more
than 60 seconds, error code F00 POWER FAIL is displayed. Press U to silence the alarm
signal. Pressing U again drains liquid from the machine and the dirty side door goes down.
90
80
70
60
50
40
30
20
10
100
10 20 30 40 50 60 min
5
CW HW HW
(DW
option)
A
B
9
3 861
HWCW+HW
7
(C)