Preventive maintenance
4
Service instructions
5015108-00 Edition 0403
Function check
This must only be done by authorised personnel.
The machine is connected to the electricity supply and some components are live.
Instructions, cable, switch
Check that a goods placing sign has been put up on the wall behind the disinfector.
Check that the isolator switch on the wall is working and that the connecting cable is
undamaged and free from defects.
Filter and valves
Check that level switches and manual shutoff valves are working properly.
Check the filters in the supply line (see under Cleaning the filters in the supply line). Clean
when necessary.
Check all pipe couplings. Tighten and seal if necessary.
Controls
The machine is operated with the buttons on the control panel.
Check that the control buttons are working. If the buttons are working, a beep will be heard.
While a program is running check that the yellow lamp at M is lit. When the program is
complete, the green lamp at N should light up.
Insert for goods
Check that the inserts for the machine dock properly with the water outlets.
Check that washing arms (if fitted) rotate and are not clogged.
Check that the goods to be cleaned are retained in the insert.
V190
Container alarm

Table of Contents