Fault indications and troubleshooting
3
Service instructions
5015108-00Edition 0403
Fault indications
Messages appear in plain text on the display. The machine cannot be started until the fault has
been put right. The following messages can be displayed:
Dose 1 low level Empty container alarm 1: if detergent 1 is finished, an operation code is
generated. The alarm is reset automatically when the container is refilled.
Dose 2 low level Empty container alarm 2: if detergent 2 is finished, an operation code is
generated. The alarm is reset automatically when the container is refilled.
Dose 3 low level Empty container alarm 3. If detergent 3 is finished, an operation
code is generated. The alarm is reset automatically when detergent is added.
Dose 4 low level Empty container alarm 4. If detergent 4 is finished, an operation code is generated.
code is generated. The alarm is reset automatically when detergent is added.
Note:
The last process had access to detergent. A message is displayed for future processes.
Red lamp: Disinfection
not achieved
Fault message or
fault code
V1363
P02 OP-D
DOSE 1 LOW LEVEL
P01
Add detergent 1
A red lamp lights up
If the red lamp at O lights up, the process has been aborted because of a fault. Call service
personnel.
Messages