Preventive maintenance
3
Service instructions
5015108-00Edition 0403
General
The required maintenance interval will depend largely on the quality of the incoming water and
how often the machine is used. The maintenance will have to be determined in each individual
case. We recommend that the stated maintenance operations are done at the specified intervals.
We also recommend that a function check is done once or twice a year.
For repairs or adjustments, see under Repair and adjustment.
Periodic maintenance
This may only be done by authorised personnel.
The machine is connected to the electricity supply and some components are live.
Component Interval
Check Replace if Replace
Replace every
Replace
yearly necessary every year other year 1000 hours
Door seal ••
Hoses between •l
dosing pump and
detergent container
Hose between dosing ••
pump and machine
Filter in incoming ••
media
Overheat protection ••
Hose in hose pump ••
Spray arm wings ••
Sterile filter in dryer ••
Hoses to dryer ••
Non-return valve in dryer ••
Check valve in waste outlet ••
Fan to dryer ••
Ink ribbon for printer ••
Manual door: Spring
on door closing
Automatic door: Crush
protection on door closing