Repair and adjustment
17
Service instructions
5015108-00Edition 0403
6. Enter the value 0 (A) and click Set Low.
7. Set the process simulator to 20 mA.
8. Enter 160 (A) in CS 1000 and click Set High.
9. Check that the value (B) rises to 160.
10. Log off CS 1000.