Repair and adjustment
21
Service instructions
5015108-00Edition 0403
7 Check that the reading of the external thermometer stabilises. Enter the reading of the external
thermometer (A) and click Set Low.
8 Check that the value (B) rises to the set value.
9 Move the temperature sensors and the external thermometer to the oil bath. Check that the
reading of the external thermometer stabilises. Enter the reading of the external thermometer
(A) and click Set High.
10.Check that the value (B) rises to the set value.
11.Log off CS 1000.
12.Re-instate the temperature sensors.