EasyManua.ls Logo

Gilson GX-271 - Cleaning Procedures

Gilson GX-271
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Gilson GX-271 Liquid Handler User’s Guide 4-3
Maintenance
4
Cleaning
Cleaning
Cleaning the Liquid Handler
The liquid handler should be cleaned occasionally using a dry, clean cloth. Or, if necessary, use a cloth dipped in soapy water. If
liquid is accidentally spilled on the liquid handler, wipe the instrument using a dry, clean cloth.
Cleaning the Syringe
Clean a syringe when some or all of the following occurs:
• Corrosive or hazardous liquids have been pumped
• Possible back flow of liquids into the waste tubing
•Leakage
• Aspiration of samples or reagents into the syringe
To clean a syringe, follow the procedures below and use the diagram as a reference.
Remove the Syringe
The following procedures use TRILUTION® LH Software.
To remove the syringe:
1 Open a TRILUTION LH Software method with 402 Syringe Pump in the
configuration.
2Click Run to open the Application Builder.
3 Add a method to the sample list.
4 Select the Manual Control tab.
5 Select a method from the drop-down list of methods.
6Click Go. The instruments will initialize.
7 Disconnect the syringe piston from the piston operating rod by unscrewing the
piston holding screw on the underside of the rod.
8Click Lower Syringe.
9 After the syringe has been lowered, unscrew and remove the syringe from the
valve.
Clean the Syringe
After the syringe has been removed, it can be cleaned:
1 Place the syringe in a beaker containing methanol. Aspirate and dispense several
volumes of methanol through the syringe.
2 Place the syringe in a beaker containing distilled or deionized water. Aspirate and dispense several volumes of water
through the syringe.
3 Hold the syringe housing in one hand. Clean the syringe using a non-abrasive cloth dampened with alcohol. Remove the
piston and clean the piston with a non-abrasive cloth dampened with alcohol.
4 Dry the syringe and piston using a clean, lint-free cloth.
valve
valve
to
syringe
piston
piston
piston
vent
tubing
operating
rod
holding
screw
screw
Figure 4-1: 402 Syringe

Table of Contents