EasyManua.ls Logo

Gilson GX-271 - Replacing the Valve

Gilson GX-271
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Gilson GX-271 Liquid Handler User’s Guide 4-13
Maintenance
4
Replacing Parts
Piston Seal for 25 mL Syringe
1 Remove the syringe from the pump as described on page 4-3.
2 Slide the piston out of the syringe barrel.
3 Hold the piston seal firmly in one hand and unscrew the seal from the piston shaft.
4 Screw a new piston seal on the shaft.
5 Dip the piston seal in distilled water and slide the piston up and down in the syringe barrel five or six times to ensure a
smooth movement between the piston and syringe.
6 Slide the piston into the syringe barrel and complete the piston seal fitting by screwing the shaft until the piston seal turns
with the piston shaft.
7 Remount the syringe on the pump as described on page 4-4
.
Replacing the Valve
To replace a valve on the syringe pump, follow the instructions below. If necessary, refer
to the syringe pump diagram.
1 Disconnect the inlet, transfer, and vent tubing from the valve.
2 Disconnect the syringe from the syringe pump. Refer to the procedure for
replacing the syringe on page 4-9
.
3 Remove the two screws securing the valve to the syringe pump and then remove
the valve.
4 Fully tighten the valve screws to secure the replacement valve.
5 Re-install the syringe. Refer to the procedure for mounting new syringe on
page 4-9
.
6 Reconnect the inlet, transfer, and vent tubing to the newly installed valve.
valve
valve
to
syringe
piston
piston
piston
vent
tubing
operating
rod
holding
screw
screw
Figure 4-12: 402 Syringe

Table of Contents