4-7
INSTALL INSTRUMENT IN STATION & PERFORM BUMP / CALIBRATION
Concurrently, the instrument automatically starts up and
begins it’s warm-up routine, if installed in the ‘OFF’ condition.
The instrument pump runs briey before the Station pump
‘overrides’ this and automatically switches OFF the instrument
pump.
The six green LED’s on the instrument ash in one second
intervals then, when the following instrument screen is
displayed, the gas delivery commences.
Note that the condence signal is disabled during bump
testing / calibration.
Fig. 4-9 Instrument Warm-up
As the gas delivery progresses, the instrument display reects
the increase in each gas value (decrease in oxygen).