4-8
AUTO BUMP / CAL STATION - HANDBOOK
The gas delivery duration is dependent on test type, as follows:
Quick Bump Test: 15 seconds default (can be edited in
conguration setup from 15 seconds to
60 seconds)
Full Bump Test: 60 seconds (xed)
Calibration: 60 seconds (xed)
When the alarm setpoints are reached, the instrument
backlight illuminates red, as illustrated in Fig. 4-10, the six
LED’s on the instrument ash red in one second intervals and
the instrument’s audible alarm activate.
Fig. 4-10 Instrument in Alarm Condition