2-7
OPERATION
Press the Left Hand (LH) button once to abort the warm-up
routine and automatically switch OFF the instrument.
Alternatively, a configurable option is available to force the user
to switch OFF the instrument.
Further details are available from the ‘CONFIGURATION
HANDBOOK’ (Part No. 64182).
2.2.7 Select Calibration Gas
This configurable option is available to allow the user to
measure a different flammable gas from that which was
originally used to calibrate the instrument.
This action allows the instrument software to compensate and
display more accurate readings.
Fig. 2-8 Cal Gas Selection
When this option is displayed, as shown in Fig. 2.8, the gas
that was originally used to calibrate the instrument is identified
between two arrowheads.
Note: The instrument calibration certificate also identifies the
original calibration gas type.
To select a different gas type, press the Left Hand (LH) button
to step through the available options from Methane,
Propane, Butane and Pentane.