3-2
USER HANDBOOK
3.2 MANUAL BUMP OPTIONS
The PS200 series provides two bump test options, QUICK
and FULL.
The QUICK bump test validates that the alarm threshold
has been exceeded for each range.
The FULL bump test checks the response of all ranges
against set limits.
The QUICK bump test is configured by default.
3.3 INITIATING A MANUAL BUMP TEST
To initiate a manual bump test, switch the instrument ON
with a long press of the LH
button. During warm-up
‘BUMP TEST’ will flash on the top line of the display, as
indicated in Fig. 3.1.
Fig. 3.1 Bump Test mode
When the warm-up is complete, the user is prompted to
apply the test gas, as shown in Fig. 3-2.