Model 672/675 Technical Reference Manual
5-7
- or -
Press [SELECT] to display “Zero=0mVv?”. Press [ENTER] to use a 0mV/V output as the zero
reference.
- or -
Press [ENTER] to display “Keyin CurW”t. Key in the known gross weight already applied to the scale
and press [ENTER].
- or -
Press [CLR] to bypass the zeroing option.
12. The display shows “CAL OK?” suggesting that the calibration is acceptable. Accept the calibration by
pressing [ENTER] at the “CAL OK?” prompt.
- or -
Repeat the calibration by pressing [ENTER] at the “CAL OK?” prompt.
13. Once the calibration is accepted in step 5, press [ENTER] at the “ENTER=SAVE” prompt and again at
the “ENTER=EXIT” prompt to save the new calibration and exit the calibration mode.
- or -
To exit the calibration mode without saving the new calibration, press [CLR]at the “ENTER=SAVE”
prompt. Then press [ENTER]at the “ENTER=UNDO” prompt and again at the “ENTER=EXIT” prompt to
exit the calibration mode.
R
R
E
E
M
M
O
O
T
T
E
E
S
S
C
C
A
A
L
L
E
E
C
C
A
A
L
L
I
I
B
B
R
R
A
A
T
T
I
I
O
O
N
N
When more than one scale is enabled, the prompt “Keyin Scl#” appears before accessing the calibration
method selections. Key in the scale number to be calibrated and press [ENTER]. Proceed with a
calibration method as described in Calibration Methods on page 5-2.
After completing a calibration, the “Keyin Scl#” appears once again. Enter the next scale number to be
calibrated, or press [CLR] to exit the calibration mode and save the new calibration data.
C
C
A
A
L
L
I
I
B
B
R
R
A
A
T
T
I
I
O
O
N
N
U
U
N
N
I
I
T
T
S
S
It is expected that a calibration weight will be entered in terms of the selected calibration units as determined
by P150 in the setup mode. The default calibration units are displayed briefly during calibration just prior to
the “Keyin CalWT” prompt.
Alternate units may be used during calibration by pressing the [UNITS] key at the “Keyin CalWT” prompt.
This will scroll through any units that have been assigned to P131 → P134 of the setup mode for the
selected scale.
It is not possible to select lb/oz as calibration units.