ASR-2000 Series User Manual
210
Power
Active (W)
Resolution
0.1 / 1 W
Accuracy
*5
±(2 % of reading +
0.5 W)
±(2 % of reading + 1
W)
Apparent
(VA)
Resolution
0.1 / 1 VA
Accuracy*
5*6
±(2 % of reading +
0.5 VA)
±(2 % of reading + 1
VA)
Reactive
(VAR)
Resolution
0.1 / 1 VAR
Accuracy
*5*7
±(2 % of reading +
0.5 VAR)
±(2 % of reading + 1
VAR)
Load power factor
Range
0.000 to 1.000
Resolution
0.001
Load crest factor
Range
0.00 to 50.00
Resolution
0.01
Harmonic voltage
Effective value (rms)
Percent (%)
(AC-INT and 50/60
Hz only)
Range
Up to 100th order of the fundamental wave
Full Scale
175 V / 350 V, 100%
Resolution
0.1 V, 0.1%
Accuracy
*8
Up to 20th ±(0.2 % of reading + 0.5 V / 1 V)
20th to 100th ±
(0.3 % of reading + 0.5 V / 1 V)
Harmonic current
Effective value (rms)
Percent (%)
(AC-INT and 50/60
Hz only)
Range
Up to 100th order of the fundamental wave
Full Scale
5 A / 2.5 A, 100%
10 A / 5 A, 100%
Resolution
0.01 A, 0.1%
Accuracy*3
Up to 20th
±(1 % of reading +
0.1 A / 0.05 A)
20th to 100th
±(1.5 % of reading +
0.1 A / 0.05 A)
Up to 20th
±(1 % of reading +
0.2 A / 0.1 A)
20th to 100th
±(1.5 % of reading +
0.2 A / 0.1 A)
*1
The voltage display is set to RMS in AC/AC+DC mode and AVG in DC mode.
*2
AC mode: For an output voltage of 17.5 V to 175 V / 35 V to 350 V and 23 °C
± 5 °C. DC mode: For an output voltage of 25 V to 250 V / 50 V to 500 V and
23 °C ± 5.
*3
An output current in the range of 5 % to 100 % of the maximum current, and 23
°C ± 5 °C.
*4
An output current in the range of 5 % to 100 % of the maximum peak current
in AC mode, an output current in the range of 5 % to 100 % of the maximum
instantaneous current in DC mode, and 23 °C ± 5 °C. The accuracy of the
peak value is for a waveform of DC or sine wave
*5
For an output voltage of 50 V or greater, an output current in the range of 10 %
to 100 % of the maximum current, DC or an output frequency of 45 Hz to 65
Hz, and 23 °C ± 5 °C.
*6
The apparent and reactive powers are not displayed in the DC mode.
*7
The reactive power is for the load with the power factor 0.5 or lower.