Abnormal Internal 
Control 
This alarm is activated and output will be disabled 
immediately when internal control abnormality is 
detected. Beware that all operations will be 
disabled except for the power shutdown operation 
if an error occurs. 
Voltage limit protection prevents a high voltage 
from damaging the DUT. This alarm can be set by 
the user. 
Frequency limit protection prevents a high 
frequency from damaging the DUT. This alarm can 
be set by the user. 
Over current protection prevents high current from 
damaging the DUT. This alarm can be set by the 
user. 
Over temperature protection for power stage 
board. OTP is a hardware protection function. Only 
when the unit has cooled can the over temperature 
protection alarms be cleared. 
Over Power protection for power stage board. OPP 
is a software protection function that is 
corresponded with VA value. When the unit has 
loaded less then protection point that the alarm be 
cleared. 
Sense alarm. This alarm will detect if the sense 
wires have been connected to the wrong polarity. 
AC input failure. This alarm function is activated 
when a low AC input is detected. 
Fan failure. This alarm function is activated when 
the fan RPMs drop to an abnormally low level.