REMOTE CONTROL
119
Graph command
:graph
Select graphing mode / path.
: graph:func
<disc>
Set the measurement function for the graph mode.
Parameter: ls lp q cs cp d z phase rs rp x g b y rdc
Example: :graph:func lp
: graph:func?
Returns the current measurement function of the
graph mode.
Return parameter: 1 (Z), 2 (Ls), 3 (Lp), 4 (Cs), 5
(Cp), 7 (Y), 8 (G), 9 (P), 10 (Q), 11 (D), 12 (Rs), 13
(Rp), 14 (B), 15 (X), 16 (Rdc), 0 (none)
: graph:sweep
<disc>
Set the sweep mode for the graph mode.
Parameter: freq, lev
Example: (drive level) :graph:sweep lev
: graph:sweep?
Returns the current sweep mode of the graph
mode.
Return Parameter: 0(frequency), 1(drive level)
: graph:st <NR3>
Set the start frequency or level for the sweep.
Parameter: (26 Hz) 26, 2.6e1, 2.600000e+01, .026k.
Example: :graph:st 2.6e1
:graph:st?
Returns the start frequency or level of the sweep.
:graph:sp <NR3>
Set the stop frequency or level for the sweep.
Parameter: (260 Hz) 260, 2.6e2, 2.600000e+02 (.26k)
Example: :graph:sp 260
:graph:sp?
Returns the stop frequency or level of the sweep.
:graph:freq
<NR3>
Set the frequency if the sweep mode is drive level.
Parameter: (150 kHz) 150000, 1.5e5, 1.500000e+05
(1.5k)
Example: :graph:freq 150k