REMOTE CONTROL 
119 
Graph command 
 
Select graphing mode / path. 
Set the measurement function for the graph mode. 
Parameter: ls lp q cs cp d z phase rs rp x g b y rdc 
 Example: :graph:func lp 
Returns the current measurement function of the 
graph mode. 
Return parameter: 1 (Z), 2 (Ls), 3 (Lp), 4 (Cs), 5 
(Cp), 7 (Y), 8 (G), 9 (P), 10 (Q), 11 (D), 12 (Rs), 13 
(Rp), 14 (B), 15 (X), 16 (Rdc), 0 (none) 
Set the sweep mode for the graph mode.  
Parameter: freq, lev 
Example: (drive level) :graph:sweep lev 
Returns the current sweep mode of the graph 
mode.  
Return Parameter: 0(frequency), 1(drive level) 
Set the start frequency or level for the sweep. 
Parameter: (26 Hz) 26, 2.6e1, 2.600000e+01, .026k. 
 Example: :graph:st 2.6e1 
Returns the start frequency or level of the sweep. 
Set the stop frequency or level for the sweep. 
Parameter: (260 Hz) 260, 2.6e2, 2.600000e+02 (.26k) 
 Example: :graph:sp 260 
Returns the stop frequency or level of the sweep. 
Set the frequency if the sweep mode is drive level. 
Parameter: (150 kHz) 150000, 1.5e5, 1.500000e+05 
(1.5k) 
 Example: :graph:freq 150k 