APPENDIX
137
Accuracy Definition
Ae[%]=±((A+0.0000001*Zx)*Kv*Kt)
Ae[%]=±((A+0.1/Zx)*Kv*Kt)
High Impedance when D < 0.1
Ae[%] = ± (( A + 0.0000001*Zx ) * Kv * Kt)
High Impedance when D ≥ 0.1
Ae[%] = ± (((A + 0.0000001*Zx ) * Kv*Kt)*
Low Impedance when D < 0.1
Ae[%] = ± (( A + 0.1/Zx ) * Kv * Kt)
Low Impedance when D ≥ 0.1
Ae[%] = ± ((( A + 0.1/Zx ) * Kv * Kt) *
High Impedance when Qx < 0.1
Ae[%] = ± (( A + 0.0000001*Zx ) * Kv * Kt)
High Impedance when Qx ≥ 0.1
Ae[%] = ± (((A + 0.0000001*Zx ) * Kv*Kt)*
Low Impedance when Qx < 0.1
Ae[%] = ± (( A + 0.1/Zx ) * Kv * Kt)
Low Impedance when Qx ≥ 0.1
Ae[%] = ± ((( A + 0.1/Zx ) * Kv * Kt) *
± ( (Ae/100) * (1+D
2
)) when D > 0.1
The Rdc accuracy is equal to Z's accuracy at the 20Hz
measurement frequency.
± (((Qx
2
*De) / (1±Qx*De)) when (Qx*De)< 1
± ((180*|Z|Ae[%]) / (π/100))
Accuracy taken from the Accuracy chart
Measured value of unknown component
w ww . . co m
information@itm.com1.800.561.8187