EasyManua.ls Logo

GW Instek LCR-8000G Series - Accuracy Definition

GW Instek LCR-8000G Series
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPENDIX
137
Accuracy Definition
|Z|, |Y|
High Impedance
[1]
Ae[%]=±((A+0.0000001*Zx)*Kv*Kt)
Low Impedance
[2]
Ae[%]=±((A+0.1/Zx)*Kv*Kt)
L, C, X, B
High Impedance when D < 0.1
Ae[%] = ± (( A + 0.0000001*Zx ) * Kv * Kt)
High Impedance when D 0.1
Ae[%] = ± (((A + 0.0000001*Zx ) * Kv*Kt)*
2
1 D
)
Low Impedance when D < 0.1
Ae[%] = ± (( A + 0.1/Zx ) * Kv * Kt)
Low Impedance when D 0.1
Ae[%] = ± ((( A + 0.1/Zx ) * Kv * Kt) *
2
1 D
)
R, G
High Impedance when Qx < 0.1
Ae[%] = ± (( A + 0.0000001*Zx ) * Kv * Kt)
High Impedance when Qx 0.1
Ae[%] = ± (((A + 0.0000001*Zx ) * Kv*Kt)*
2
1 Q
)
Low Impedance when Qx < 0.1
Ae[%] = ± (( A + 0.1/Zx ) * Kv * Kt)
Low Impedance when Qx 0.1
Ae[%] = ± ((( A + 0.1/Zx ) * Kv * Kt) *
2
1 Q
)
D
± ( Ae/100) when D 0.1
± ( (Ae/100) * (1+D
2
)) when D > 0.1
Rdc
The Rdc accuracy is equal to Z's accuracy at the 20Hz
measurement frequency.
Q
± (((Qx
2
*De) / (1±Qx*De)) when (Qx*De)< 1
Ɵ
± ((180*|Z|Ae[%]) / (π/100))
Convention
A
Accuracy taken from the Accuracy chart
Zx
Measured value of unknown component
Kv
Test Voltage factor
Level
Kv
Level
Kv
1.250
1.2
0.078
2
0.625
1
0.039
2.5
0.313
1.2
0.02
5
0.156
1.5
0.010
10
Kt
Temperature factor
Temperature
Kt
Temperature
Kt
8-18°C
2
28-35°C
2
18-28°C
1
w ww . . co m
information@itm.com1.800.561.8187

Table of Contents