LCR-8000G Series User Manual 
24 
Fixture Connection 
Fixture structure 
 
The standard fixture is a four-wire type with a 
common terminal for screen connection. The outer 
terminals (Hforce and Lforce) provide the current 
and the inner terminals (Hsense and Lsense) 
measures the potential. 
LFORCE LSENSE HSENSE HFORCE
+–
GND
(Optional)
Carries the signal current source. 
Connected to the + side of the device 
under test. 
Together with Lsense, monitors the 
Potential. Connected to the + side of 
the device under test. 
Together with Hsense, monitors the 
Potential. Connected to the – side of 
the device under test. 
Accepts the signal current return. 
Connected to the – side of the device 
under test. 
If the test component has a large 
metal area NOT connected to either 
of the terminals, connect the GND 
clip to minimize noise level. 