LCR-8000G Series User Manual
24
Fixture Connection
Fixture structure
Background
The standard fixture is a four-wire type with a
common terminal for screen connection. The outer
terminals (Hforce and Lforce) provide the current
and the inner terminals (Hsense and Lsense)
measures the potential.
Diagram
LFORCE LSENSE HSENSE HFORCE
+
GND
(Optional)
Description
HFORCE
Carries the signal current source.
Connected to the + side of the device
under test.
HSENSE
Together with Lsense, monitors the
Potential. Connected to the + side of
the device under test.
LSENSE
Together with Hsense, monitors the
Potential. Connected to the side of
the device under test.
LFORCE
Accepts the signal current return.
Connected to the side of the device
under test.
GND
If the test component has a large
metal area NOT connected to either
of the terminals, connect the GND
clip to minimize noise level.