PASS-FAIL MODE
77
Edit program step
How to edit
parameter
For selecting parameters, press F1
(Prog) repeatedly.
F 1
For entering values, use the numerical and unit
keys.
Example:
0.5kHz
k
Enter
Backspace
All clear
Clear
Increase
Decrease
How to move
cursor
To move the cursor to the editing point,
use the arrow keys: up, down, left, and
right.
Note: The cursor cannot move into
blank columns or steps.
Panel operation
1. Create new
step
To create a new step in a program,
press F1 (Prog) at Func column “OFF”.
The step becomes active and the
function changes to Ls. Up to 30 steps
are available in a program.
F 1
Step
Func
OFF
01
Step
Func
Ls
01
2. Select item
(function)
Move the cursor to Func column, press
F1 (Prog) repeatedly. The measurement
item (function) changes in the following
order.
F 1
Ls Lp Q Cs Cp D Z θ Rs Rp
X G B Y Rdc Ls