LCR-8000G Series User Manual
76
2. Enter the new program name using the arrow
keys.
Move cursor
(Left/Right keys)
JKLMNOPQRS
Enter the letter
(Down key)
program name: N_
Delete one letter
(Up key)
program name: _
3. Press the Enter key to confirm the
file name. To quit the new program
mode, press the Clear key.


4. A blank program with the entered name
appears.
MULTI STEP MODE
Delete
F 1
F 2
F 3
F 4
F 5
Step
PROGRAM: New
Set
Func
Freq
Volt
Bias
Spd
Hi
Lo
Dly
01 02 03
OFF OFF OFF
Prog
F 6
Copy
Save
File
RUN
5. Press F1 (Prog). Step 01 becomes
active and changes to Ls
measurement mode. For further
editing details, see the next page.
F 1
Step
Func
OFF
01
Step
Func
Ls
01