BASIC MEASUREMENT
53
Set measurement frequency
*This setting does not apply to Rdc measurement.
Background
The measurement frequency, together with the
measurement voltage, defines the electrical
condition of each measurement item. Make sure
the appropriate frequency is selected, according to
the component characteristics.
Panel operation
1. Press the Left/Right key repeatedly
to move the cursor to Frequency.
2.00 195.00 kHzVac
2. Enter the frequency using the numerical keys.
Range
20Hz ~ 1MHz (LCR-8101G)
20Hz ~5MHz (LCR-8105G)
20Hz ~10MHz (LCR-8110G)
1.2kHz
k
Enter
1MHz
M
Enter
Backspace
All clear
Clear
Increase
Decrease
When the entered value does not fit in the
range, the LCR-8000G automatically selects the
nearest value.
Nearest Available
When the wrong unit (such as Ω) is entered,
the value is cancelled.
Unit Mismatched