GETTING STARTED 
29 
10b. Select 
Repetitive 
measurement 
Press the Sing/Rep key to select 
Repetitive (automatic trigger) 
measurement. Press the Left/Right arrow 
key and move the cursor to Speed. Press 
the Up/Down key to select the speed. 
11a. Select 
Absolute 
measurement 
Press F5 (Abs/%/∆) to select Abs. Press 
the Left/Right key to move the cursor to 
Lo (Low limit). Use the numerical and 
unit keys to set the Low limit. Repeat this 
for Hi (Hi limit) as well. 
11b. Select 
Percentage 
measurement 
Press F5 (Abs/%/∆) to select %. Press the 
Left/Right key to move the cursor to the 
Nominal value. Use the numerical and 
unit keys to set the numerical level. Then 
move the cursor to Lo (Low limit) and set 
the percentage. Repeat this for Hi (Hi 
limit) as well. To save the latest 
measurement result as Nominal, press F6 
(Save Nom). 
11c. Select Delta 
measurement 
Press F5 (Abs/%/∆) to select ∆. Press the 
Left/Right key to move the cursor to the 
Nominal value. Use the numerical and 
unit keys to set the numerical level. Then 
move the cursor to Lo (Low limit) and set 
the Lo value difference. Repeat this for Hi 
(Hi limit) as well. To save the latest 
measurement result as Nominal, press F6 
(Save Nom). 
To hide the drive Voltage/Current, press 
the Code key, type 80, then press Enter. 
Set the Range (internal setting) to Auto, 
use the Left/Right key to move the cursor 
and Up/Down key to change the setting. 