APPENDIX
243
Accuracy of Setting
*1
±(0.6 % of set + 1.4 % of f.s
*2
) + Vin
2*3
/500k(3.24M)
*1 It is not applied for the condition of the parallel operation.
*2 M range applies to the full scale of H range.
*3 Vin = Input terminal voltage of electronic load.
Soft Start
Operation Mode
CC, CR and CP
Selectable Time Range
1- 200 ms/Res: 1ms
Time Accuracy
±(30%of set + 100us)
Remote Sensing
Voltage that can be Compensated
2V for a single line
Protection Function
Model
PEL-3021(H)
PEL-3041(H)
PEL-3111(H)
Overvoltage protection(OVP)
Turns off the load at 110% of the rated voltage
Overcurrent protection(OCP)
0.03-38.5A
(0.0060A-9.6252A)
0.06A-77A
(0.0120A-19.2504A)
0.2A-231A
(0.050A-57.750A)
or 110% of the maximum current of each range
Load off or limit selectable
Overpower protection(OPP)
0.1W - 192.5W
0.3W - 385W
1W - 1155W
or 110% of the maximum power of each range
Load off or limit selectable
Overheat protection(OTP)
Turns off the load when the heat sink temperature reaches 105 °C
(PEL-3211H:115°C )
Undervoltage protection(UVP)
Turns off the load when detected. Can be set in the range of
0V(0.1V) to 150V(840V) or Off.