PEL-3000(H) User Manual
68
Operation
1. Make sure the load is off.
2. Press
Main
> Configure[F5] > Other[F2] and
set the Dyna. Level setting.
Range:
Value, Percent
Display:
Percent Setting
Mode
CP
I Range
H 8.75A
V Range
L 80V
Function
Dynamic
Configure
0.000
V
0.0000
A
0.00
W
Timer1 0.025
Level 50.0
Set 1.00
LOADRS232
Fine
01/Oct/2018
ms
%
W
Set level
% from the Set value
Example
Level
Time
Set
Level
Timer1 Timer2 Timer1
Dynamic Mode:
Dyna. Level = Percent
Select the Switching Time Configuration for Dynamic Mode
Description
The switching time for dynamic mode can be
configured to switch between two preset on-times
(Timer1, Timer2) or by setting a switching
frequency and duty cycle.
Operation
1. Press
Main
> Configure[F5] > Other[F2] and
set the Dyna. Time setting.
Range:
T1/T2, Freq. Duty