74
Von Latch
Description
When Von Latch is set to ON, the load will
continue to sink current after being “latched”, even
if the voltage drops below the Von Voltage
threshold level.
When Von Latch is set to OFF, the load will turn
off when the voltage drops below the Von Voltage
threshold level.
By default Von Latch is set to OFF.
Time
Von Latch = OFF
Von
V
Time
I
PEL-3000
V out
Load on
Load off Load on
Time
Von Latch = ON
Von
V
Time
I
PEL-3000
V out
Load on
Operation
1. Press
Main
> Configure[F5] > Other[F2] and
set the Von Latch setting.
Range:
Von Latch: OFF, ON