OPERATION
65
Basic Configuration
The basic configuration settings are the common configuration
settings that are used for each operating mode. After selecting a
basic operating mode (CC, CR, CV or CP mode), the slew rate,
switching function, response rate and other common parameters
should be configured.
Select the Switching Function
Description
The PEL-3000(H) has two switching function, static
and dynamic. The switching function allows the
PEL-3000(H) to switch between two preset levels.
Static mode can only switch between the two levels
manually, while Dynamic mode switches between
each level automatically based on a timer.
Static mode: A Value, B Value
Dynamic mode: Level1, Level2
When the unit is set to static mode, only one value
(A Value or B Value) can be active at a time. The
active value is shown in the Operation Status
Panel.
Function
Static
Configure
0.00
W
LOADRS232
A Value
Fine
W
W
Active
setting
When the unit is set to dynamic mode, the unit will
switch between Level1 and Level2 based on the
Timer1 and Timer2 parameters, shown below.