Calibration Data
DECIMAL PLACES 2
DECIMAL PLACES setting defines the number of decimal
places (0, 1, 2 or 3) system displays in the reaction results and
in the relevant system menus.
TOC CALIBRATION 1
TOC CALIBRATION 2
TOC CALIBRATION 3
TIC CALIBRATION 1
TIC CALIBRATION 2
TIC CALIBRATION 3
TN CALIBRATION 1
TN CALIBRATION 2
TN CALIBRATION 3
Calibration menus are used to install the calibration data and
thus to calibrate the BioTector in factory. TOC/TIC/TN
Calibration 1 displays the calibration curves for range 1,
Calibration 2 for range 2 etc.
The first column shows the number of calibration points, the
second point [%] shows the calibration factors automatically
calculated from the calibration data, the third column
STANDARD shows the standard solution concentrations and the
fourth column shows the un-calibrated result (e.g. mgC/l, mgN/l
etc.) measured and calculated from the CO
2
analyzer and Dual
Cell Photometer readings.
System calibration curves should not be modified by the user on
site unless a recalibration of the system is required at a different
analysis range.
Sequence Program
Average Program
LOG AVERAGE NO
LOG AVERAGE function enables (YES) or disables (NO by
default) the 24-hour averaging of the reaction results in the
reaction archive.
The reaction results obtained from manual samples are not
included in the averaging calculations.
AVERAGE UPDATE 00:00
If the LOG AVERAGE above is enabled (YES), the system
calculates a 24 hours average of the reaction results and
displays it in the reaction archive for each stream at the time
programmed for AVERAGE UPDATE (00:00, HR:MIN, by
default).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B7000 and is the answer not in the manual?

Hach BioTector B7000 Specifications

General IconGeneral
BrandHach
ModelBioTector B7000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals