9.3 BioTector Notification Conditions
The following Notification conditions do not activate the BioTector stop sequence, leave the 4-20mA signal
un-changed and do not activate the fault relay.
85_REAGENTS LOW
(can also be programmed as
a warning condition)
BioTector has calculated that the
reagent containers are running low.
Change reagents, and reset
reagent monitor in Install New
Reagents menu.
Confirm reagent volume/container
size input values are correct in
Reagents Monitor menu.
Power applied to the BioTector or
system reboot after the processor
watchdog timeout.
Self accepting notification.
No action is required.
This message is logged into the fault
archives when “RESET SERVICE
COUNTER” function is selected in
Diagnostics, Service menu.
Self accepting notification.
No action is required.
This message is logged into the fault
archives when external device sends a
sample fault input signal to the
BioTector.
See Fault Setup section for details.
Check the external sample liquid
level and sampling system for the
sample channel.
Check the external sample
monitoring device and the external
input signal wiring.