EasyManuals Logo

Hach BioTector B7000 User Manual

Hach BioTector B7000
175 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
Page 146
9.3 BioTector Notification Conditions
The following Notification conditions do not activate the BioTector stop sequence, leave the 4-20mA signal
un-changed and do not activate the fault relay.
NOTIFICATION
CONDITION
CAUSE/REMEDY
85_REAGENTS LOW
(can also be programmed as
a warning condition)
BioTector has calculated that the
reagent containers are running low.
Change reagents, and reset
reagent monitor in Install New
Reagents menu.
Confirm reagent volume/container
size input values are correct in
Reagents Monitor menu.
86_POWER UP
Power applied to the BioTector or
system reboot after the processor
watchdog timeout.
Self accepting notification.
No action is required.
87_SERVICE TIME RESET
This message is logged into the fault
archives when RESET SERVICE
COUNTER” function is selected in
Diagnostics, Service menu.
Self accepting notification.
No action is required.
122 SAMPLE FAULT 1
This message is logged into the fault
archives when external device sends a
sample fault input signal to the
BioTector.
See Fault Setup section for details.
Check the external sample liquid
level and sampling system for the
sample channel.
Check the external sample
monitoring device and the external
input signal wiring.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B7000 and is the answer not in the manual?

Hach BioTector B7000 Specifications

General IconGeneral
BrandHach
ModelBioTector B7000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals