The meaning of the abbreviations used in the downloaded reaction data in both standard and engineering
modes (see Print Mode in Data Program, Printer, PC, MMC/SD Card menus) are as follows:
Standard Mode:
TIC & TOC Analysis:
TIME The time the reaction started.
DATE The date the reaction started.
S1:2 Reaction type and range.
TICmgC/l The calibrated TIC value in mgC/l.
TOCmgC/l The calibrated TOC value in mgC/l (TOC represents NPOC).
TNmgN/l The calibrated TN value in mgN/l.
COD/BODmgO/l The calculated COD and/or BOD value in mgO/l (if activated in COD/BOD Program menu).
TC Analysis:
TIME The time the reaction started.
DATE The date the reaction started.
S1:2 Reaction type and range.
TCmgC/l The calibrated TC value in mgC/l (TC represents TIC + NPOC + POC).
TNmgN/l The calibrated TN value in mgN/l.
VOC Analysis:
TIME The time the reaction started.
DATE The date the reaction started.
S1:2 Reaction type and range.
TCmgC/l The calibrated TC value in mgC/l (TC represents TIC + NPOC + POC).
TICmgC/l The calibrated TIC value in mgC/l.
TOCmgC/l The calculated TOC value in mgC/l (TOC is calculated as TC TIC).
VOCmgC/l The calculated VOC value in mgC/l (VOC is calculated as TC TIC NPOC).
TNmgN/l The calibrated TN value in mgN/l.
TC - TIC Analysis:
TIME The time the reaction started.
DATE The date the reaction started.
S1:2 Reaction type and range.
TCmgC/l The calibrated TC value in mgC/l.
TICmgC/l The calibrated TIC value in mgC/l.
TOCmgC/l The calculated TOC value in mgC/l (TOC is calculated as TC TIC).
TNmgN/l The calibrated TN value in mgN/l.