Page 140
FAULT
CONDITION
CAUSE/REMEDY
18_LIQUID LEAK DET
BioTector’s cabinet liquid leak detector
is activated.
Check for a liquid leak in the
BioTector.
Check BioTector liquid leak
detector.
19_DCP LIQ. LEAK DET
BioTector’s DCP module liquid leak
detector is activated.
Check for a liquid leak in the
BioTector DCP module.
Check DCP module liquid leak
detector.
20_NO REAGENTS
(can also be programmed as
a warning or notification
condition)
BioTector has calculated that the
reagent containers should be empty.
Change reagents, and reset
reagent monitor in Install New
Reagents menu.
Confirm reagent volume/container
size input values are correct in
Reagents Monitor menu.