Page #164 background imageLoading...
Page #164 background image
Page 164
As in Stream Multiplex Mode, in Full Multiplex Mode, the Stream ID channel also operates in discrete levels
as shown in table A3.2 below.
Table A3.2 4-20mA STREAM ID channel output levels.
STREAM ID CHANNEL
Output Level (mA)
CHANGE
4
STM 1
5
STM 2
6
STM 3
7
STM 4
8
STM 5
9
STM 6
10
MAN 1
11
MAN 2
12
MAN 3
13
MAN 4
14
MAN 5
15
MAN 6
16
CAL
17
CAL Z
18
CAL S
19
RESERVED
20
In Full Multiplex Mode, the result type channel also operates in discrete levels. This is shown in table A3.3
below.
Table A3.3 RESULT TYPE channel output levels.
RESULT TYPE CHANNEL
Output Level (mA)
NOT_DEFINED
4
TIC
5
TOC
6
TC
7
VOC
8
TN
9
RESERVED
10
RESERVED
11
RESERVED
12
RESERVED
13
RESERVED
14
RESERVED
15
RESERVED
16
RESERVED
17
RESERVED
18
RESERVED
19
RESERVED
20
As can be seen in table A3.3 above, the Result Type channel has some levels reserved for future
development as the detection capability of BioTector systems is extended.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B7000 and is the answer not in the manual?

Hach BioTector B7000 Specifications

General IconGeneral
BrandHach
ModelBioTector B7000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals