If BioTector is built as a TOC TN analyzer, the analysis results of the relevant parameters (e.g. TN, COD,
BOD etc.) are also displayed on one or more screens.
These screen give information on:
- The Reaction Start time.
- The Reaction Type, for example a TN STREAM 2 reaction.
- The Reaction Phase, for example if the reaction is currently in the SAMPLE PURGE phase etc.
- The Range the BioTector is using to carry out its analysis.
- The current Reaction Time for the corresponding reaction.
- The Reaction Duration for the overall duration of the corresponding reaction.
Reaction Graph
The Reaction Graph screen gives information on the current reaction in progress, and allows the user to
monitor the progression of the reaction. This screen gives information on:
- The current atmospheric pressure, measured in kPa.
- The un-calibrated data from the reaction, for example TICmgu or TOCmgu without any compensation
for atmospheric pressure.
- The height of the CO
2
peaks in each phase of the reaction.
- The current MFC flow in l/h.
- The temperature in the analyzer.
- The CO
2
instantaneous value (CO2i) and the CO
2
zero value (CO2z) for the reaction.
- The reaction time.