MONITORING RANGES: TOC / TN
Standard Range 0-10mg/l up to 0-20,000mg/l
Ultra High Range 0-10mg/l up to 0-40,000mg/l
Up to 3 ranges configurable for each component within each range band detailed above.
A wide combination of TOC and TN monitoring ranges, including higher ranges, are available on
request.
Exceedance Tracking: Full Exceedance Tracking to Maximum Range
Range Selection: Automatic or Manual Range Selection
Repeatability: 3% of reading or 0.3mg/l TOC whichever is greater,
3% of reading or 0.2mg/l TN whichever is greater,
with Automatic TOC TN Range Selection feature
Detection Limit: 0.6mg/l TOC with Automatic Range Selection
0.4mg/l TN with Automatic Range Selection
Sodium Chloride Interference:
TOC
All Ranges None
TN**
0 10 mgN/l None below 5% w/v
0 100 mgN/l None below 6% w/v
0 1000 mgN/l None below 30% w/v
0 10000 mgN/l None below 30% w/v
0 100000 mgN/l None below 30% w/v
** The sodium chloride interference figures are range dependent and are typically defined for 0.5mm path length TN measuring cell
assuming that the TOC:TN ratio is 1.
SAMPLE & ENVIRONMENTAL CONDITIONS
Sample Volume: Up to 8.0ml
Sample Inlet Pressure: Typically ambient (for applications with high sample pressure, optional
sampling systems are available)
Drain Pressure: Typically ambient (for applications with high drain pressure, optional systems
are available)
Sample Inlet Temperature: 2°C 60°C (36°F - 140°F)
Sample Flow Rate: Minimum 100ml per sample
Sample Particle Size: Up to 2 mm, soft particulates
Ambient Temperature: 5°C 40°C (41°F - 104°F)
Air conditioning and heating options are available.
Humidity: 5% - 85%, non-condensing
Ingress Protection: IP44
Optional IP54 with air purge
System Sound: < 60 dBa
The manufacturer has a continuous research and development program. Specifications may
therefore be changed without notice. For specification updates, please contact the
manufacturer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach BioTector B7000 and is the answer not in the manual?

Hach BioTector B7000 Specifications

General IconGeneral
BrandHach
ModelBioTector B7000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals