Download BioTector “All Data” in text format into the MMC/SD card or into a PC/Laptop using
the SEND ALL DATA function in Data Output menu to record all changes made in the system
configuration.
_____
Go to Start Stop menu and start the BioTector. When the BioTector is running online,
carefully observe the first two or three reactions and confirm that the CO
2
peaks are correct.