2. Hang the analyzer and make it vertical as follows:
a. Hang the analyzer mounting brackets on the fasteners. Push the analyzer against the wall.
b. Put the level behind the door latches on the surface that is parallel to the wall. Make sure that
the analyzer is vertical within ±10 mm (0.39 in.). Use shims as necessary.
c. Tighten the fasteners until the analyzer cannot be lifted.