10. Remove the analyzer from the wall as follows:
a. Loosen the fasteners for the analyzer until the heads of the fasteners are 0.6 mm (¼ in.) from
the wall.
b. Lift the analyzer mounting brackets up and off of the fasteners.
11. Put the analyzer back in the original packaging.
Clean the instrument
N O T I C E
Never use cleaning agents such as turpentine, acetone or similar products to clean the instrument including the
display and accessories.
Clean the exterior of the instrument with a moist cloth and a mild soap solution.
Clean spills
C A U T I O N
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.
1. Obey all facility safety protocols for spill control.
2. Discard the waste according to applicable regulations.
Troubleshooting
Remote indicator lights and sounds
The analyzer status and the chlorine status are identified with sounds and colored lights on the
remote indicator in the treatment room. Refer to Table 10 and Table 11. Similar alert and alarm
sounds occur at the remote indicator and the analyzer.
When the battery life of the remote indicator is low, the remote indicator makes a continuous sound.
Replace the battery when the continuous sound is heard. Refer to Replace the remote indicator
battery on page 60.
Note: Alarm and alert settings cannot be changed. Sound volumes and tones cannot be changed.