Connect the remote indicator cable to the analyzer
Install the female blade terminal on the end of the remote indicator cable. Then, connect the remote
indicator cable to the remote indicator connector.
1. Use the CAT5 cable stripper to remove 38 mm (1.5 in.) of the outer cable shielding from the
remote indicator cable.
2. Untwist each pair of wires. Use the wire strippers to remove 7 mm (0.25 in.) of plastic from each
pair of twisted wires.
3. Install the female blade terminal on the end of the remote indicator cable as follows:
a. Twist the silver stranded (shield) wires together.
b. Put the silver stranded wires into the female blade terminal.
c. Use a crimper tool to close the female blade terminal on the silver stranded wires.
d. Gently pull the silver stranded wires to make sure that the wires are held securely in the
female blade terminal.
If step 3 is not completed correctly, cut off the loose wires and do steps 1 and 3 again.