Replace the remote indicator battery
D A N G E R
Multiple hazards. Do not operate the remote indicator without a battery. The battery is a safety feature.
When the battery life of the remote indicator is low, the remote indicator makes a continuous sound.
Replace the battery yearly or when the continuous sound is heard.
Items supplied by the user:
Lithium coin battery CR2032 (3 V) Cross-slotted (Phillips) screwdriver
Pre-requisite: Remove power from the analyzer as follows:
1. Push menu. Select Power>shut down.
2. Wait for the analyzer to shut down. When the display is off, disconnect the power cord from the
electrical outlet.
Note: For user safety, the analyzer will automatically go back to idle mode if the power cord is not disconnected
in 5 minutes or less after the display goes off.
Note: When power is removed from the analyzer for more than 10 minutes, a power-loss beep is heard at the
remote indicator for approximately 5 minutes.
Replace the remote indicator battery
1. Remove the cover of the remote indicator as follows:
a. Loosen the captive screw at the top of the remote indicator.
b. Remove the cover of the remote indicator.
2. Remove and discard the battery.
60
English