Start guided installation
1. Push menu. Select Maintenance>Installation (scroll up). Complete the steps that show on the
display. The display screens are shown for reference in Appendix A: Screens—Install the
colorimeter on page 74.
2. Wait for the analyzer to shut down. When the display is off, disconnect the power cord from the
electrical outlet.
Note: For user safety, the analyzer will automatically go back to idle mode if the power cord is not disconnected
in 5 minutes or less after the display goes off.
Note: When power is removed from the analyzer for more than 10 minutes, a power-loss beep is heard at the
remote indicator for approximately 5 minutes.
3. Remove the pump clamp from the reagent delivery module.