Hach CM130 - Guided Setup and Component Installation

Hach CM130
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Start guided installation
1. Push menu. Select Maintenance>Installation (scroll up). Complete the steps that show on the
display. The display screens are shown for reference in Appendix A: Screens—Install the
colorimeter on page 74.
2. Wait for the analyzer to shut down. When the display is off, disconnect the power cord from the
electrical outlet.
Note: For user safety, the analyzer will automatically go back to idle mode if the power cord is not disconnected
in 5 minutes or less after the display goes off.
Note: When power is removed from the analyzer for more than 10 minutes, a power-loss beep is heard at the
remote indicator for approximately 5 minutes.
3. Remove the pump clamp from the reagent delivery module.
English 49