7. Negative pressure therapy
The activities in this chapter may be carried out only by a physician or a qualified person�
Two therapy modes are available:
> Continuous mode
> Intermittent mode
Continuous mode
The factory setting is 125mmHg in continuous mode�
Generally the most recent settings are always stored�
7.1 Setting the negative pressure
7.1.1 Continuous mode
Intermittent
Continuous
Main menu
> Tapping on the
key increases the negative pressure in
5mmHg steps�
> Tapping on the
key decreases the negative pressure in
5mmHg steps�
7.1.2 Intermittent mode
Intermittent
Continuous
Intermittent mode
In contrast to continuous mode, which operates with a constant negative
pressure, intermittent mode allows therapy with changing pressure
intervals�
Switching intermittent mode on
> Tap on the
key�
Intermittent mode is activated.
The surrounding key border turns white.
The factory setting for intermittent mode is 125mmHg for 5minutes and
20mmHg for 2minutes.
The most recent setting is always stored�
> Tap on the
area�
The settings menu for intermittent mode is displayed.
> Tap on the value that is to be changed�
The affected field turns white.
> Tap on the
or the key to set the desired value�
> Confirm the entry by pressing
�
The activated intermittent mode is displayed.
Intermittent
Continuous
Change-over range for settings