28
Prioritisation of the warning messages is in accordance with the following table, in decreasing priority�
Automatic switch-o
The device will automatically
switch o in a few minutes.
Connect the mains plug
Automatic switch-off
If no power supply was connected despite repeated “Battery discharged!”
warning messages, the negative pressure unit switches itself off
automatically 1minute after the warning message�
IMPORTANT!
The negative pressure unit can be switched on again
after the power supply cable has been reconnected and
therefore the rechargeable battery is recharged�
Blocked tube
1. Check tube for kinks
2. Check connections
3. Check exudate canister
Blocked tube
The warning message “Blocked tube” appears when the negative pressure
unit identifies a blockage in the exudate canister or in the tubing system�
The delay until this warning message condition is determined is
3�5-8�5minutes (+/-5seconds)�
WARNING!
The device is not able to identify a blockage for negative
pressure settings below 50mmHg� For this reason, the
wound dressing must be checked frequently for correct
compression of the foam�
This warning message can have the following causes:
• A kink in the tubing system
> Position the tube such that no kink can arise�
• Blockage at the connection points
Check all connection points for possible blockages or incorrect
connections�
• Check the exudate canister
> Tap on the
key�
The warning message is suppressed for 5minutes.
IMPORTANT!
If the fault cannot be rectified by way of the above-
mentioned measures, the exudate canister should
be changed�
IFU_4095080_0302884_190824_EN.indd 28IFU_4095080_0302884_190824_EN.indd 28 19.08.24 12:3119.08.24 12:31