28
29
Prioritisation of the warning messages is in accordance with the following table, in decreasing priority
Automatic switch-o
The device will automatically
switch o in a few minutes.
Connect the mains plug
Automatic switch-off
If no power supply was connected despite repeated “Battery discharged!”
warning messages, the negative pressure unit switches itself off
automatically 1minute after the warning message�
IMPORTANT!
The negative pressure unit can be switched on again
after the power supply cable has been reconnected and
therefore the rechargeable battery is recharged
Blocked tube
1. Check tube for kinks
2. Check connections
3. Check exudate canister
Blocked tube
The warning message “Blocked tube” appears when the negative pressure
unit identifies a blockage in the exudate canister or in the tubing system�
The delay until this warning message condition is determined is
3�5-8�5minutes (+/-5seconds)�
WARNING!
The device is not able to identify a blockage for negative
pressure settings below 50mmHg� For this reason, the
wound dressing must be checked frequently for correct
compression of the foam�
This warning message can have the following causes:
A kink in the tubing system
> Position the tube such that no kink can arise�
Blockage at the connection points
Check all connection points for possible blockages or incorrect
connections�
Check the exudate canister
> Tap on the
key�
The warning message is suppressed for 5minutes.
IMPORTANT!
If the fault cannot be rectified by way of the above-
mentioned measures, the exudate canister should
be changed�
IFU_4095080_0302884_190824_EN.indd 28IFU_4095080_0302884_190824_EN.indd 28 19.08.24 12:3119.08.24 12:31