Hartmann Vivano Tec Pro - Electrical Safety; General Precautions

Hartmann Vivano Tec Pro
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
7
potential contact of intestinal content with the wound, enteric fistula needs to be surgically separated, following local
guidelines or established surgical practices�
Spinal cord injuries with the development of autonomic hyperreflexia
Discontinue the negative pressure wound therapy if the patient has spinal cord injuries with the development of
autonomic hyperreflexia�
Magnetic Resonance Imaging
This device is not considered as MRI safe and must not be used in close proximity to an MRI unit�
Defibrillation
The VivanoTecPro unit must be disconnected if resuscitation of the patient using a defibrillator is necessary
Hyperbaric oxygen therapy (HBO)
The VivanoTecPro unit must be disconnected for patients undergoing hyperbaric oxygen therapy, as its use is considered
a potential fire hazard�
External heat sources
Keep the negative pressure unit away from sources of heat and flames
Electrical safety
Before connecting the device, check that the mains voltage and mains frequency indicated on the device correspond to
the values for the supply network� The connecting cable and the accessories must be checked for damage before use of
the negative pressure unit
IMPORTANT: Damaged cables must be replaced immediately
NOTE: Only medical and undamaged mains connections must be used� Multiple socket outlets or extension cables may
not be used�
IMPORTANT: The patient must not take a bath or shower while the VivanoTecPro negative pressure unit is in place�
The therapy should be interrupted for this purpose only after consultation with the attending physician�
IMPORTANT: Never touch the mains plug or the power supply with wet hands and never touch the power supply cable or
the DC input and the patient at the same time
IMPORTANT: No alterations may be made to the unit or the supplied power supply
Flammable or explosive gases and/or liquids
The negative pressure unit must not be used in the presence of readily flammable or explosive gases and/or liquids�
IMPORTANT: The negative pressure unit is not designed for use within areas at risk of explosion and oxygen-enriched
areas� Areas at risk of explosion can occur through the use of flammable anaesthetic agents, (or mixtures with air, oxygen
or nitrous oxide), skin cleansers and skin disinfectants�
1.4 General precautions
Please pay attention to the following precautions:
Damaged, expired or contaminated product
Do not use any component of the VivanoSystem in case of damage, expiration or any suspicion of contamination� It may
cause overall decline in the therapeutic efficiency, wound contamination and/or infection�
Single use only
All disposable components of the VivanoSystem are meant for single use only� Reusing a single-use medical device is
dangerous� Reprocessing devices, in order to reuse them, may seriously damage their integrity and their performance�
Information available on request