29
Leakage
1. Check wound dressing
2. Check connections and exudate canister
Leakage
This error message appears in the case of a leakage in the system which
cannot be compensated for by the negative pressure unit�
The delay until this warning message condition is determined is 2minutes
(+/-5seconds)�
> Check the wound dressing for possible leakage�
> Check all connections for leakage�
> Check if the exudate canister is firmly connected to the negative
pressure unit�
If it is not possible to rectify the leakage through these measures, it might
be advisable to replace the wound dressing and the tubing system�
> Tap on the
key�
The warning message is suppressed for 5minutes.
Technical defect
Please send back for service.
Technical defect
The negative pressure unit is no longer functioning correctly and is
possibly damaged�
> The negative pressure unit is no longer in an operational condition�
Return it to your specialist dealer or to PAULHARTMANNAG for
checking and repair�
Exudate canister full
Please change exudate canister
Exudate canister full
The delay until this warning message condition is determined is
2- 47seconds (+/-1second)�
This error message can be triggered by the following causes:
• The exudate canister is full�
> Change the exudate canister�
• Blocked bacteria filter in the exudate canister� If the bacteria filter is
wetted by the exudate, it will block�
> Change the exudate canister�
> Tap on the
key�
The warning message is suppressed for 5minutes.
IMPORTANT!
To avoid a blockage in the filter, the negative pressure
unit must always remain in an upright position and must
not tilt�
The therapy must be restarted after the exudate canister
has been changed (see chapter Starting the therapy)�