32
Leakage
1. Check wound dressing
2. Check connections and exudate canister
Leakage
This warning message is displayed if the system detects a leak which
cannot be compensated for by the negative pressure unit�
> Notify your attending physician or nursing staff immediately�
> Tap on the
key�
The warning message is suppressed for 5minutes
Technical defect
Please send back for service.
Technical defect
The negative pressure unit is no longer functioning correctly and is
possibly damaged�
Warning cannot be removed – Continues display�
> Notify your attending physician or nursing staff immediately�
Exudate canister full
Please change exudate canister
Exudate canister full
This warning message is displayed when the canister is full�
> Notify your attending physician or nursing staff immediately� The
exudate canister must be changed without delay so that the therapy is
not interrupted�
> Tap on the
key�
The warning message is suppressed for 5minutes.
Battery discharged
Connect the mains plug
Battery discharged
This message is displayed when the remaining operating time is less than
one hour (see chapter “Preparing the negative pressure unit for use”�
> Push the plug of the power supply into the socket on the negative
pressure unit�
> Connect the power supply to the supplied country-specific power
supply cable�
> Plug the mains plug into the mains socket�
> Tap on the
key�
The warning message is suppressed for 15minutes.
The therapy can meanwhile be continued unimpeded.
Impact
Please send back for service.
Impact
The negative pressure unit is no longer functioning correctly and is
possibly damaged�
> Notify your attending physician or nursing staff immediately�
> Tap on the
key�
The warning message will not appear again until the negative pressure unit
is switched on again.