39
14. Technical data
VivanoTecPro
Essential performance
The essential performance characteristic of the negative pressure unit is the
generation and maintenance of the set negative pressure (+ 7%) or the generation
of a warning message conforming with EN 60601-1-8 (for further information refer
to Technical Data Sheet)�
VivanoTecPro power supply Model: AFM45US24C2-XE1047
Input: 100-240 V ac, 1�1 A, 50-60Hz
Output: 24 V dc 2�01 A
Manufacturer: XP Power Limited
Model: AKM45US24C2-XZ1579
Input: 100-240 V ac, 1�1 A, 50-60Hz
Output: 24 V dc 2�0 A 48W
Manufacturer: XP Power Limited
VivanoTecPro power supply
cable
Length: 5m
VivanoTecPro The device is either powered by external power supply or internally powered (input:
14,40 - 14,52 V dc, 2 A)
Operation time Charging time approx� 2h, operation up to 16h, (80mmHg and 0�4l / min), leakage
and capacity display on the touchscreen, uninterrupted operation on mains power
supply
Negative pressure regulation max� 200mmHg, min� 20mmHg in steps of 5mmHg, requires a connected exudate
canister for a stable negative pressure control
Display Touchscreen
Operating modes Continuous and intermittent
Data memory Internal memory for therapy data: 1Gb
Intermittent mode Time interval: 2-10minutes in each case
Upper pressure value: 40-200mmHg
Lower pressure value: 20-80mmHg
The preset minimum pressure difference between upper pressure and lower pressure
value is 20mmHg�
Standard values for the upper pressure range: 5min� 125mmHg
Standard values for the lower pressure range: 2min� 20mmHg
Continuous mode Standard value 125mmHg; pressure range: 20-200mmHg
IFU_4095080_0302884_190824_EN.indd 39IFU_4095080_0302884_190824_EN.indd 39 19.08.24 12:3119.08.24 12:31